Relative Viscosity of Plasma - Evaluation of a Simple Technique

Shoaib Tauheed  ( Department of Physiology, Dow Medical College, Karachi. )
Shaheen Shoaib  ( Department of Physiology, Dow Medical College, Karachi. )

Introduction

The measurement of serum/plasma viscosity is valuable in different diseases. The laboratory apparatus for measuring viscosity is not widely available and viscometry is not done at all in this country. Keeping in view its importance, a simple disposable syringe method of Leonard1 was used in this study for measurement of relative viscosity of serum or plasma.

Patients, Methods and Results

Relative viscosity (RV) of plasma was measured in 155 adult males, 30 clinically healthy non-smokers, 40 healthy smokers, 35 hypertensives and 50 patients with ischaemic heart disease. Blood was drawn from subjects in the resting state in the morning with minimum stasis and anticoagulated with EDTA in concentration of 1.5 mg/ml blood. Plasma was obtained after centnfuging the blood at 3000 rpm.
Relative viscosity of plasma was measured by the method of Leonard1. His disposable system consists of a standard 2 ml plastic syringe, internal diameter0.8 cmanda 19 G serum needle. The barrel of the syringe in the disposable system was filled by drawing up undiluted sample through the needle starting withthe plunger at the 1 ml mark. Taken in this fashion, the sample level was just over 2 ml mark at the moment that the plunger was withdrawn completely from the banel of the syringe with the syringe hand held in a vertical position, the run thmugh time was started when the bottom of the meniscus passed the 2 ml mark of the barrel. The RVofthe sample was derived by dividing the run through time for the plasma against distilled water at the same temperature. The procedure was done at 23°C. Fibrinogenwas determinedby kit method Sigma with sodium citrate as anticoagulant.

Table I shows that RV of plasma in 155 cases is 1.98±0.02 and the correlation between RV and fibrinogen is strong with rvalue 0.75.

Table II shows the relative viscosity of plasma was 1.62±0.021 (range was 1.52 - 1.80) in healthy controls, 1.929±0.017 in smokers, 2.087±0.029 in hypertensives and 2.172±0.036 in patients with ischaemic heart disease. RV was significantly higher (P>0.001) in smokers and patients with hypertension and ischaemic heart disease.

Comments

The viscosity ofabulk liquid is its intrinsic resistance to flow, which arises because of internal friction between its molecular and particulate components. Issac Newton in 1686 hypothesized that shear rate in a fluid was directly propor­tional to shear stress. Newtonian fluids like plasma maintain a constant viscosity at any flow velocity whereas non-Newto­nian fluids like whole blood change viscosity at different flow velocities2. The normal plasma viscosity is the same for both sexes and for all ages in normal subjects3. The incidence of false positive and false negative results are significantly lower. It can also be done in postal specimen and it is independent of haematocrit and other red cell factors. It correlates better with disease activity, there is less interference by steroid or salicylate therapy, it can also be used for diagnosis of plasma hyperviscosity syndrome4. An increase in any plasma protein fraction including albumin increases plasma viscosity5. It also is sensitive to acute phase reactant protein in the monitoring of clinical progress6 as well as changes in chronic disease2. The increase in viscosity above water..viscosity is largely because of fibrinogen and is due to its high molecular weight and assymmetry3. As observed earlier plasma/viscosity was higher in healthy smokers7,8 and patients with hypertension and ischaemic heart disease9-12. Leonard1 performed the disposable syringe procedure on 49 serum samples and compared it with Oswald’s viscometer. A direct correlation of RV determined by Leonard’s method and fibrinogen, a stxongfactordetennining the viscosity was observed in this study. This study has also established the reference values for clinically healthy subjects in our community but there is a need for extensive work especially by using advanced viscometers according to the recommendation of ICSH13, so that the viscosity of plasma as well as that of whole blood can be measured at different shear rates. Measurement of relative viscosity of plasma is a simple rapidly carried out test with good precision which is clearly feasible and reproducible as an out as well as inpatients test and may be of value in assessing risk factors for various diseases and in diagnosing hyperviscosity states.

References

1. Leonard, R. C. F. Simple technique for measuring serum or plasma viscosity with disposable apparatus. Br. Med. J., 1981 ;283: 1154-55.
2. Lowe, G. D. 0. Blood rheology and hyperviscosity syndrome. Bailliere\\\'s clinical haematology. London, Bailliere Tindall, 1987, pp. 598-605.
3. Harkness, J. A new instrument for the measurement ofplasma viscosity. Lancet, 1963;ii:280-81.
4  Harkness, 3. Theviscosity ofhuman blood plasma, its measurement in health and disease. Biorheology, 1971 ;8: 171-93.
5. Eastham, R. D. and Morgan. E. H. Plasma viscosity in clinical laboratory practices. J. Med. Lab. Tech., 1965;22:70-73.
6. Hutchison, R. M. and Eastham, R. D. A comparison of eiythrocyte sedimenta­tion rate and plasma viscosity in detecting changes in plasma proteins. J. Clin. Path., 1977,30:345-49.
7. Galea, G. and Davidson, R. J. L. Haematological and haemorheological changes associated with cigarette smoking. 3. Clin. Pathol., 1985;38:978-84.
8. Ernst, E., Matrai, A., Schomolzl, C. H. et al. Dose effect relationship between smoking and blood rheology. Br. 3. Haematol., 1987;65:485-87.
9. Tibblin, G., Bergentz, S. E., Bjure. J. et al. Haematocrit, plasma protein, plasma volume and viscosity in early hypertensive disease. Am. Heart J., 1966;72:165-76.
10. Letcher, R. L., Chien, S., Pickering, T. J. et al. Direct relationship between blood pressure and blood viscosity in normal and hypertensive subjects. Role of fibrinogen and concentration Am. 3. Med., 1981;70:1195-1202.
11. Lowe, 0. D. 0.. Drummond, M. M., Lorimer, A. R. et al. Relation between extent of coronary artery disease and blood viscosity. Br. Med. J., 1980;1:673-74.
12. Meade, T. kW., Brozovick, M., Chakarbarty, R. et al. Haemostatic function and cardiovascular death. Principle results of Northwick Park Heart Study. Lancet, 1986;ii:533-37.
13. International committee for standardization in haematology. Recommendations for a selected method for measurement of plasma viscosity. J. Clin. Pathol., 1984,37:1147-54.

Journal of the Pakistan Medical Association has agreed to receive and publish manuscripts in accordance with the principles of the following committees: