Maqsood Ahmad Burq ( Department of Ophthalmology, Aga Khan University Hospital, Karachi Pakistan )
Jehanzeb Khan ( Department of Ophthalmology, Aga Khan University Hospital, Karachi Pakistan )
Sidra Zafar ( Department of Ophthalmology, Aga Khan University Hospital, Karachi Pakistan )
Baila Shakaib ( Department of Ophthalmology, Aga Khan University Hospital, Karachi Pakistan )
Zafar Iqbal ( Department of Ophthalmology, Al Shifa Hospital, Karachi, Pakistan. )
July 2022, Volume 72, Issue 7
Research Article
Abstract
Objective: To analyse the effect of pupil dilation on intraocular lens instrument IOLMaster 500 biometric measurements, and to determine the effect of these measurements on intraocular lens power calculations in Asian eyes.
Methods: The prospective study was conducted at the Aga Khan University Hospital, Karachi, between January and April 2017, and comprised all patients scheduled for cataract surgery who underwent scanning with IOLMaster 500. For each patient, pre-dilation and post-dilation measurements were taken. The intraocular lens power was determined through Sanders/Retzlaff/Kraff Theoretical, Holladay, and Hoffer Q formulae. Data was analysed using SPSS 24.
Results: There were 276 eyes of 138 participants who had a mean age of 59.7±11.1 years. Anterior chamber depth changed significantly with pupil dilation (p=0.001). No significant changes were observed in the axial length (p=0.410), keratometry measurements (p=0.931), and intraocular lens power calculations (p>0.05).
Conclusions: The change in anterior chamber depth, though significant, was perhaps clinically non-significant.
Introduction
Cataract surgery has transformed in the past few decades. Modern cataract surgery is safe, with much faster postoperative recovery periods. Not surprisingly, the procedure is now one of the most performed worldwide.1,2 However, one of the biggest challenges that still remain with phacoemulsification is accurate calculation of intraocular lens (IOL) power.
Precise and accurate IOL power calculations are central to a successful cataract procedure and postoperative patient satisfaction. IOL power is affected by three variables; keratometry (K) measurements, anterior chamber depth (ACD) and axial length (AL). A mistake in calculation of any of these variables may thus be a source of error.3 Patients scheduled for cataract surgery routinely undergo pupil dilation as part of a comprehensive examination. Mydriasis may produce changes in the corneal shape and, hence, corneal power. Accommodation changes and fixation may also be more difficult in the presence of mydriasis, which can introduce errors in AL measurement.4
A few studies have been done to examine the effect of pupil dilation on IOLMaster measurements.4-8 However, most of these studies have been conducted in the Western population and similar literature from Asian nations is sparse. The current study was planned to fill the gap by determining the accuracy of IOLMaster biometric measures pre- and post-dilation in an Asian setting.
Patients and Methods
The prospective study was conducted in the outpatient ophthalmology clinics at the Aga Khan University Hospital (AKUH), Karachi, between January and April 2017. After approval from the institutional ethics review committee, the sample size was calculated using openepi.com.9 We used an assumed mean AL of 23.4±1.36mm in both measurements,5 the correlation between both measurements to be 0.8, and a difference <0.2mm between both measurements to be clinically insignificant, with a statistical significance of 5% and a statistical power of 80%.
All patients aged >18 years scheduled for cataract surgery who underwent scanning with IOLMaster 500 and willing to participate were included. Those with pre-existing ocular pathology, such as glaucoma, uveitis and retinal disease, and previous ocular surgery, were excluded.
After taking informed consent, biometric measurements were done using IOLMaster 500 (Carl Zeiss, Germany) due to its better accuracy in biometric measurements in comparison with ultrasound.10,11 It is a noncontact optical instrument that uses infrared light (λ=780nm) of short coherence for measuring AL and ACD, and for performing K. IOL power was calculated using the built-in software.
All patients underwent consecutive biometric measurements, pre- and post-pupillary dilation. ACD was defined as the distance from the anterior corneal surface to the anterior lens surface, and AL as the distance from the anterior corneal surface to the retinal pigment epithelium. Central corneal thickness (CCT) was the distance from the anterior corneal surface to the posterior corneal surface.
Pupillary dilation was achieved using 1% tropicamide and 10% phenylephrine due to reduced problems with glare accomodation.12 The second set of measurements was taken after 45 minutes to ensure maximum mydriasis. All patients also underwent a complete ophthalmological examination, including slit-lamp fundoscopy. IOL power was predicted based on the Hoffer Q, Holladay, and the Sanders/Retzlaff/Kraff Theoretical (SRK/T) formulas. The accuracy of these formulas has been excellent.13-19 Each eye acted as its own control, and the measurements were paired. All readings were taken in a standardised environment under mesopic lightening.
Data was analysed using SPSS 24. The variations were analysed using paired t tests. P<0.05 was considered significant.
Results
Of the 158 subjects initially enrolled, 138(87.34%) were included, translating into 276 eyes. The mean age of the participants was 59.7±11.1 years. No significant differences were observed in measurements for AL, K1, K2 and mean K after pupillary dilation (p>0.05), while ACD significantly increased post-dilation (Table 1).
Mean change in lens power was not significant (Table 2).
Hypermetropic eyes were defined as AL <22.5mm, and emmetropic eyes had AL range 22.5-24.5mm. Myopic eye was defined as AL >24.5mm. No significant changes were observed in IOL power post-dilation for any AL group using all 3 formulae (Table 3).
Of the 276 studied eyes in the current sample, 193(70%), 165(60%) and 158(57%) eyes had no change in the calculated IOL before and after dilation using SRK/T, Holladay and Hoffer Q formulae, respectively. In the remaining cases where IOL measurements were different pre-dilation and post-dilation, the change was within 0.5 deviation (D) in 71(27.5%) eyes when SRK/T formula was used. It was within 1D and >1D in 6(2.2%) cases each. Using Holladay and Hoffer Q, the difference in lens power was within 0.5D in 97(35%) and 96(34.8%) eyes. The difference was within 1D in 8(2.9%) eyes when Holladay formula was used, and in 15(5.4%) eyes with Hoffer Q. The change was >1D in only 6(2.2%) eyes with both formulae (Table 4).
Discussion
There was a significant change in ACD with cycloplegia, while the remaining measures, including IOL power, did not exhibit any statistically significant change in the current study.
Previous studies evaluating the effect of pupil dilation have reported similar outcomes. AL has been found to remain unaffected in all the reported studies.4,5 K measurements, in the form of K2 and mean K significantly varied with mydriasis in only one study.5 Despite changes in K readings, the study5 did not observe significant changes in IOL power. Prospective studies6,7 observed significantly increased ACD post-cycloplegia, a finding that was also observed in the current study. The device used by Villalobos et al20 was an optical low-coherence reflectometry (OLCR) optical biometer, instead of the gold standard partial coherence-based biometry device. Deepening of the ACD could partly be explained by the lens flattening and moving more posteriorly with accommodative paralysis.
With respect to IOL power, significant differences in lens power have been reported only with the Haigis formula21 which is a fourth-generation formula that calculates effective lens position, taking into account the ACD. However, till date no significant differences in lens power have been observed when using the SRK/T and Holladay formulae as the prediction models.
The current study’s findings related to pre- and post-dilation measurements using SRK/T, Holladay and Hoffer Q formulae were similar to results reported earlier.4
Although SRK/T has been previously analysed, to the best of our knowledge, this is the first study evaluating Holladay and Hoffer Q in an Asian sample.
The current study, however, has limitations. Even though AKUH caters to an ethnically diverse group of patients, the study was only done at a single centre. Multi-centre studies involving subjects from different ethnicities and countries would strengthen the findings. Besides, the study did not include the Haigis formula to predict IOL power calculation which also takes into account the ACD while making calculations.
Conclusion
There were no significant changes in AL, K measurements and IOL power calculations post-mydriasis in all three prediction models. However, there was a significant difference in ACD readings. The clinical significance of this difference, however, is debatable and, in the light of the results of other studies conducted earlier, it is perhaps non-significant.
Disclaimer: None.
Conflict of Interest: None.
Source of Funding: None.
References
1. Dick HB, Schultz T. A review of laser-assisted versus traditional phacoemulsification cataract surgery. Ophthalmol Ther 2017; 6: 7-18.
2. Linebarger EJ, Hardten DR, Shah GK, Lindstrom RL. Phacoemulsification and modern cataract surgery. Surv Ophthalmol 1999; 44: 123-47.
3. Olsen T. Calculation of intraocular lens power: a review. Acta Ophthalmol 2007; 85: 472-85.
4. Adler G, Shahar J, Kesner R, Rosenfeld E, Fischer N, Loewenstein A, et al. Effect of pupil size on biometry measurements using the IOLMaster. Am J Ophthalmol 2015; 159: 940-4.
5. Heatley CJ, Whitefield LA, Hugkulstone CE. Effect of pupil dilation on the accuracy of the IOLMaster. J Cataract Refract Surg 2002; 28:1993-6.
6. Rodriguez-Raton A, Jimenez-Alvarez M, Arteche-Limousin L, Mediavilla-Peña E, Larrucea-Martinez I. Effect of pupil dilation on biometry measurements with partial coherence interferometry and its effect on IOL power formula calculation. Eur J Ophthalmol 2015; 25: 309-14.
7. Khambhiphant B, Sasiwilasagorn S, Chatbunchachai N, Pongpirul K. Effect of pupillary dilation on Haigis formula-calculated intraocular lens power measurement by using optical biometry. Clin Ophthalmol 2016; 10: 1405-10.
8. Huang J, McAlinden C, Su B, Pesudovs K, Feng Y, Hua Y, et al. The effect of cycloplegia on the lenstar and the IOL Master biometry. Optom Vis Sci 2012; 89: 1691-6.
9. Sullivan KM, Dean A Soe MM. Open Epi: a web based epidemiologic and statistical calculator for public health. Public Health Rep 2009; 124: 471-4.
10. Olsen T. Improved accuracy of intraocular lens power calculation with the Zeiss IOLMaster. Acta Ophthalmol Scand 2007; 85: 84-7.
11. Ueda T, Taketani F, Ota T, Hara Y. Impact of nuclear cataract density on postoperative refractive outcome: IOL Master versus ultrasound. Ophthalmologica 2007; 221: 384-7.
12. Eyeson-Annan ML, Hirst LW, Battistutta D, Green A. Comparative pupil dilation using phenylephrine alone or in combination with tropicamide. Ophthalmology 1998; 105: 726-32.
13. Retzlaff J. A new intraocular lens calculation formula. J Am Intraocul Implant Soc 1980; 6: 148-52.
14. Sanders DR, Kraff MC. Improvement of intraocular lens power calculation using empirical data. J Am Intraocul Implant Soc 1980; 6: 263-7.
15. Retzlaff J. Posterior chamber implant power calculation: regression formulas. J Am Intraocul Implant Soc 1980; 6: 268-70.
16. Sanders D, Retzlaff J, Kraff M, Kratz R, Gills J, Levine R, et al. Comparison of the accuracy of the Binkhorst, Colenbrander, and SRK™ implant power prediction formulas. J Am Intraocul Implant Soc 1981; 7: 337-40.
17. Sanders DR, Retzlaff J, Kraff MC. Comparison of empirically derived and theoretical aphakic refraction formulas. Arc Ophthalmol 1983 101: 965-7.
18. Thompson JT, Maumenee AE, Baker CC. A new posterior chambe intraocular lens formula for axial myopes. Ophthalmology 1984; 91 484-8.
19. Sanders DR, Retzlaff J, Kraff MC. Comparison of the SRK II formul and other second-generation formulas. J Cataract Refract Surg 1988 14: 136-41.
20. Saw VP, Paula CD, Shweikh Y, Deshmukh RR, Restori M. Outcomes o the Haigis-L formula for calculating intraocular lens power afte myopic and hyperopic laser refractive surgery in a tertiary teachin hospital. Investigative Ophthalmology & Visual Science 2017; 58 1150.
21. Wong CW, Yuen L, Tseng P, Han DC. Outcomes of the Haigis-L formul for calculating intraocular lens power in Asian eyes after refractiv surgery. J Cataract Refract Surg 2015; 41: 607-12.
Related Articles
Journal of the Pakistan Medical Association has agreed to receive and publish manuscripts in accordance with the principles of the following committees: