Percutaneous Balloon Valvuloplasty Inpatients with Pulmonary Valve Stenosis: A Single Center Experiment

Alireza Ahmadi,Mohammadreza Sabri  ( Department of Pediatrics, Pediatric Cardiology,Faculty of Medicine, and Child Health Promotion Research Center, Isfahan University of Medical Sciences, Isfahan, Iran. )

March, 2012

Abstract

Background: Pulmonary valve stenosis (PVS) is one of the most important acyanotic congenital heart diseases in children. Over the last years, balloon pulmonary valvuloplasty (BPV) has become the treatment of choice for the relief of moderate to severe valvular pulmonic stenosis in all age groups. The aim of current study was to report our regional survey on the short term outcomes of BPV among patients with PVS treated in Isfahan Hearth center.
Methods: Thirtyseven patients with moderate and severe PS based on echocardiography, aged between 0.6 and 16 years, underwent catheterization and angiography. Balloon pulmonary valvuloplasty was performed for thirty four patients. Right ventricle and pulmonary artery pressure and pressure gradient was measured before and after valvuloplasty.
Results: The pulmonary valvuloplasty was successful in 32 (85%) of 34 attempts. Immediately after dilation, right ventricle and pulmonary artery pressure and pressure gradient decreased from 73.0 ± 48.0to 38.4 ± 23.1 mmHg, from 35.2 ± 22.1 to 20.7 ± 9.7 mmHg and from 79.0 ± 32.1 to 33.4 ± 19.9 mmHg, respectively (p <0.001).There was no major complication. Two patients with significant subvalvar PS referred to surgeon for infundibular resection.
Conclusion: BPV was an effective procedure to treat PVS in patients with moderate to severe stenosis. It effectively reduced right ventricle-pulmonary artery systolic pressure gradient with low complication.
Keywords: Pulmonary valve stenosis, Balloon pulmonary valvuloplasty, Treatment (JPMA 62: S-58; 2012).


Introduction

Cardiac malformation with an estimate prevalence of 5 per 1000 live birth is the most common congenital abnormality. Valvular defects are considered the most common cardiac malformation from which 8-12% is related to pulmonary valve stenosis (PVS).It is considered the most benign valvular lesion.1
Pulmonary valve stenosis primarily results from a maldevelopment of the pulmonic valve tissue and the distal portion of the bulbuscordis. The stenosis is classified as valvular, subvalvular, or supravalvular. Stenosis may be presented in a typical modification inmost of the cases (80-90%) and as the effect of dysplastic changes in 10-20% of cases.2,3
Pulmonic valvular stenosis represents the most frequent cause of right ventricular outflow obstruction, and its related adverse outcomes such as decreased cardiac output, right ventricular hypertrophy, early congestive heart failure (CHF) and cyanosis are directly associated with the right ventricular systolic pressure gradient. Accordingly, it is classified to mild and moderate-to-severe forms. Management of the disease depends on the degree of stenosis.4,5
Over the last few years, balloon pulmonary valvuloplastyhavebecome the treatment of choice for the relief of moderate to severe valvular pulmonic stenosis in neonates, infants, children, and adults.6 The procedure was first described by Kan et al. It has similar outcomes comparable to surgical valvotomy and is a less invasive, more effective and safe method for relief of right ventricular obstruction in affected children, so surgical pulmonary valvotomy was replaced with this method.7,8
Since then many studies worldwide have investigated the short- and long-term effects of BPV, reduction of the pulmonary valve pressure gradient, normalization of hemodynamic disorders, the need of second intervention, age of treated patients, its complications in patients with PVS and methods for improving the procedure and its outcomes. According to their reports, though the exact indication of BPV is not defined yet but should probably be similar to those used for surgical valvotomy; it is indicated in patients with moderate to severe PVS.9 The aim of current study was to report our regional survey on the short-term outcomes of BPV among patients with PVS.


Methods

In this cross-sectional study, children with moderate and severe pulmonary stenosis aged 0.6-16 years diagnosed in Chamran Hospital, Isfahan-Iran, during 2007-2009 were selected. Those with previous history of surgical valvuloplasty, children with systemic diseases that may affect cardiac function and those that required balloon valvuloplasty as palliative treatment for cyanosis were excluded. Pediatric cardiologist selected the patients who were candidate for BPV. Selected patients underwent cardiac catheterization and angiography and BPV was performed. Right ventricle and pulmonary artery pressure and pressure gradient was measured before and after BPV.
PVS diagnosis and BPV:
PVS diagnosed based on physical examination, ECG, chest radiograph and echocardiographic, hemodynamic and angiocardiographic findings. Before catheterization, the pulmonary artery flow, peak systolic pulmonary artery flow, the size of pulmonary valve annulus for detecting hypoplastic valve, poststenotic dilatation of the pulmonary artery, right ventricular size, wall thickness, and function was measured. The assessments was done using a 3.5-MHz, quantitative, range-gated, two-dimensional, pulsed, echocardiographic Doppler scanner with fast Fourier transform spectral output and a 2.5-MHz phased array with pulsed or continuous-mode Doppler. Color Doppler evaluation in conjunction with 2D echocardiography was used to confirm the clinical diagnosis and quantitating the degree of PVS.
Peak Doppler velocity was used to predict the catheter measured peak to peak gradients across the pulmonary valve. The peak instantaneous Doppler gradient calculated by a modified Bernoulli equation P4V2 (P: the pressure gradient and V:the peak Doppler flow velocity in the main pulmonary artery).10 The right ventricular peak systolic pressure (RVP) estimated by a modified Bernoulli equation:
RVP = 4V2+ERAP (V: the peak tricuspid regurgitant jet velocity and ERAP:the estimated right atrial pressure).
Doppler measurements were performed in studied children in resting state and those who were anxious sedated mildly. Different morphology of the pulmonary valve was defined in four categories as follows.
Typical: mild to moderately thickened leaflets with evidence of commissural fusion and normal annular dimensions
Dysplastic: severely thickened leaflets with evidence of nodular hyperplasia or annular hypoplasia
Combined: dysplastic valve morphology as defined but with additional evidence of commissural fusion
Complex: other morphologies including primarily post-surgical valvotomy valves and valves involved as part of complex congenital heart lesions
Mild, moderate and severe PVS was defined according to valve gradient as follows.
Mild PS: a valve gradient <35 to 40 mmHg Moderate PS: the gradient >40 mmHg Severe PS: a gradient >60 to 70 mmHg
After selecting patients with moderate to severe PVS, hemodynamic cardiac catheterization and angiography was percutaneously performed to confirm the clinical impression and to consider balloon dilatation of the pulmonary valve. Right ventricle and pulmonary artery pressure and peak to peak pressure gradient between RV and PA was measured before and after valvuloplasty.
Studied patients were sedated with intermittent doses of midazolam (0.05 to 0.1 mg/kg IV) and/or fentanyl (0.5 to 1.0 microgram/kg IV) or ketamine before performing balloon pulmonary valvuloplasty. During procedure, ECG monitoring, blood pressure and arterial oxygen saturation measurement was performed by means of a pulse oximeter.
The percutaneous femoral venous route was the most preferred entry site for balloon pulmonary valvuloplasty. A 5 to 7 Fr sheath was inserted into the vein regarding the age and size of the patient as well as the anticipated size of the balloon dilatation catheter. After determining the severity and location of stenosis, an exchange guide wire was introduced through an end-hole catheter and positioned in the distal left pulmonary artery. A balloon that was larger than 20-40% of pulmonary valve annulus size measured by angiography was used and positioned over a guide wire with the valve at its midpoint. As the balloon was inflated, a waist from the stenotic valve was initially observed and disappeared at full inflation (Figure).


Immediately after BPV, reduction of the gradient of transvalvular systolic pressure, reduction of end-diastolic RV pressure, changes of the systolic pressure in the pulmonary artery, presence/intensification of subvalvular stenosis of the pulmonary artery, appearance of insufficiency of the pulmonary valve, or other complications were evaluated.
An effective intervention was defined as one which decreased the stenosis by at least 50% of the initial value.11,12 Outcome was classified according to the RV-PA pressure gradient immediately after dilation: Successful if the RV-PA gradient dropped to <36 mmHg and unsuccessful if the RV-PA gradient remained >36 mmHg.13Patients were transferred to the intensive care unit until stabilization.
Statistical Analysis:
Obtained data were analyzed using SPSS (version 16) software and Student\'s t-test with 0.05 level of statistical significance was employed. Calculations were made using the STATISTICA statistical program.


Results

In this study, 37 patients (11 men and 26 women) were studied.Mean age of studied patients was 4.7 ± 2.8 years with a range of 0.6-16 years. According to echocardiographic findings, typical, dysplastic and complex morphology of the pulmonary valve was presented in 82%, 13% and 5% of patients, respectively. According to echocardiographic findings, 6 (16%) of patients had moderate PVS and 31(84%) had severe PVS. Cardiac catheterization findings indicated that mild, moderate and severe PVS was presented in 3(8.1%), 14(37.8%) and 20(54.1%) of patients. Therefore, 34 patients with moderate and severe PVS enrolled.
Among studied population, co-existing cardiac malformations were reported as follows; patent foramen ovale (PFO), small atrial septal defect (ASD), subvalvularpulmonary stenosis (PS), mild aortic stenosis (AS), small ventricular septal defect (VSD), Noonan and hypertrophic cardiomyopathy (HCMP)in 16, 3, 3, 2, 1, 1 and 1 patients, respectively.
A good outcome, defined as a residual catheter gradient 36 mmHg, was achieved in 77% of the total group and in 85% of those with typical valve morphology. In contrast, 65% of patients with dysplastic pulmonary valve had a suboptimal outcome. The balloon diameter to pulmonary valve annulus ratio was 1.28 ± 0.2.There was no major complication. Two patients with significant subvalvar PS referred to surgeon for infundibular resection.
According to our findings, we could estimate the size of balloon for BPV based on the age of the patient using this formula:Size of balloon=13.5 +0.89x(x=age of patient).The pulmonary valvuloplasty was accomplished in 32 (94%) of 34 attempts.Mean ± SD of right ventricular systolic pressure (RVSP), pulmonary artery pressure (PAP)and the transvalvular peak to peak systolic gradient (DP) before and after intervention is presented in Table.


Discussion

The regional findings of BPV in patients with moderate to severe PVS were investigated. Percutaneous balloon pulmonary valvuloplasty was effective and safe for the treatment of PVS. Additionally, we could estimate the size of balloon for the procedure regarding the age of the patient.
As mentioned earlier, since the first description of balloon pulmonary valvuloplasty in 1982, balloon dilatation of valvular pulmonic stenosis is now the standard treatment for patients with PVS of all ages and all sizes especially the moderate to severe forms, and evidences from different regions confirms its utility.13-15
In this study, most of the patients had typical morphology of PVS and those with dysplastic morphology had severe form of PVS, which was similar to those reported by Piotr Werynski et al. in Poland. In their study, 92.4% of patients had typical PVS and those with dysplastic form were associated with moderate to severe forms.16
The immediate dilation success rate in our study was 77%(85% in typical forms).The rate was reported to be 88.9% in the study of Liu et al. in China,17 81.5% in another study in China,18 and 90% in the study of Jaing et al.19 Our findings and the results of previous studies indicated the effectiveness of BPV in the treatment of PVS. The success rate in our study was lower than mentioned ones because 18% of the studied patients had dysplastic or complex morphology. Our results were similar to the findings of Handoka et al. in Egypt with a success rate of 78%.20 Previous studies showed that factors such as infundibular stenosis and pulmonary valve dysplasia were associated with a lower success rate.21
In this study, mean of RVSP, PAP and the transvalvular peak to peak systolic gradient (dP) after intervention decreased significantly in patients with moderate to severe PVS. It was in line with the findings of Jaing et al. in Taiwan,19 Liu et al. in China,17 and Piotr Werynski in Poland.16
Handokaet al. in a similar study among 23 patients with severe PVS reported a significant reduction of the right ventricular systolic pressure, RV-PA systolic pressure gradient and elevation of the pulmonary artery systolic pressure. Moreover, they identified the predictors of successful pulmonary balloon valvuloplasty and concluded that the immediate post-PBV gradient is the most important predictor in this field.20
The balloon diameter to pulmonary valve annulus ratio was 1.28 ± 0.2, which was appropriate for this procedure. Studies in this field indicated that a balloon/annulus ratio of 1.2-1.5 is considered the best fordilatation of the pulmonary valve for appropriate immediate and intermediate outcomes.22 Moreover, Rao et al. has recently reported that balloon/annulus ratio of 1.2-1.25 may produce optimal results.11 In addition, we obtained a formula regarding the size of balloon and patients age, it would be useful for selection of appropriate balloon based on age.
In current study, we did not observe any complication during and after BPV. Complication rate reported by similar study was different. Overall, the rate of complication of this procedure is low. According to the report of Valvuloplasty and Angioplasty of Congenital Anomalies (VACA) registry, the rate of severe, moderate and mild complication of BPV was 0.6%, 1.3% and 2.6%, respectively. The complications are as follows; injury of vessel, tricuspid valve, annulus or the cuspid of the pulmonary valve, disruption of the pulmonary trunk, perforation of the RV outflow tract, closure of RV outflow tract, arrhythmias, cardiac tamponade and sudden cardiac arrest.23
The limitation of current study was small sample size of patients and short period of study, i.e. we did not follow up our studied patients for long-term to evaluate the outcome of BPV. In conclusion, BPV was an effective procedure for treatment of PVS in patients with moderate to severe stenosis, as it effectively reduced RV-PA systolic pressure gradient with low complication.


References

1. Armstrong EJ, Bischoff J. Heart valve development: endothelial cell signaling and differentiation. Circ Res 2004; 95: 459-70.
2. Freedom RM, Benson L. Congenital pulmonary stenosis and isolated congenital pulmonary insufficiency. The natural and modified history of congenital heart disease.New York: Blackwell Pub 2004; pp 107-718.
3. Libby P, Bonow RO, Mann DL, Zipes DP. Braunwald\'s Heart Disease; A Textbook of Cardiovascular Medicine. 8thed. Philadelphia: WB. Saunders Inc; 2007.
4. Nishimura RA, Pieroni DR, Bierman FZ, Colan SD, Kaufman S, Sanders SP, et al. Second natural history study of congenital heart defects. Pulmonary stenosis: echocardiography. Circulation 1993; 87(2 Suppl): I73-9.
5. Crawford M. Pulmonic stenosis. In: Crawford M. Cardiology. 3rded.St. Louis, MO: Mosby Inc; 2009.
6. Cheragh H, Hassan M, Hafizullah M, Gul AM. Outcome of balloon pulmonic valvuloplasty with 18 months follow up. J Ayub Med Coll Abbottabad 2009; 21: 95-9.
7. Kan JS, White RI Jr, Mitchell SE, Gardner TJ. Percutaneous balloon valvuloplasty: a new method for treating congenital pulmonary-valve stenosis. N Engl J Med 1982; 307: 540-2.
8. Schmaltz AA, Bein G, Grävinghoff L, Hagel K, Hentrich F, Hofstetter R, et al. Balloon valvuloplasty of pulmonary stenosis in infants and children--co-operative study of the German Society of Pediatric Cardiology. Eur Heart J 1989; 10: 967-71.
9. Rao PS. Balloon pulmonary valvuloplasty: a review. Clin Cardiol 1989; 12: 55-74.
10. Goldberg SJ, Allen HD, Marx GR, DonnersteinR. Doppler echocardiography. 1sted. Philadelphia: Lea & Febiger; 1988; pp 26-7, 113-37.
11. Rao PS. Percutaneous balloon pulmonary valvuloplasty: State of the art. Catheter Cardiovasc Interv 2007; 69: 747-63.
12. Moore P, Lock JE. Catheter intervention, balloon angioplasty, experimental studies, technology and methodology. In: Lock JE, Keane JF, Perry SB, Editors. Diagnostic and interventional catheterization in congenital heart disease. 2nded. Boston, MA: Kluwer Academic Publishers; 2000; pp 140-2.
13. Hatem DM, Castro I, Haertel JC, Rossi RI, Zielinsky P, Leboute FC, et al. Short- and long- term results of percutaneous balloon valvuloplasty in pulmonary valve stenosis. Arq Bras Cardiol 2004; 82: 221-7.
14. Abu Haweleh A, Hakim F. Balloon pulmonary valvuloplasty in children:Jordanian experience. J Saudi Heart Assoc 2003; 15: 31-4.
15. Akçurin G, Kahramanyol O, Atakan C. Intermediate-term follow-up results of pulmonary balloon valvuloplasty in children. Turk J Pediatr 2000; 42: 126-31.
16. Werynski P, Rudzinski A, Król-Jawien W, Kuzma J.Percutaneous balloon valvuloplasty for the treatment of pulmonary valve stenosis in children - a single centre experience. Kardiol Pol 2009; 67: 369-75.
17. Liu F, Wu L, Huang GY, Zhang LF, Qi CH, Ma XJ, Liang XC. Percutaneous balloon valvuloplasty for severe and critical pulmonary valve stenosis in infants under six months. Zhonghua Yi Xue Za Zhi 2009; 89: 3253-6.
18. Yu ZX, Ma YT, Yang YN, Huang D, Ma X, Fu ZY, Liu F. Outcome of percutaneous balloon pulmonary valvuloplasty for patients with pulmonary valve stenosis. Zhonghua Xin Xue Guan Bing Za Zhi 2009; 37: 1006-9.
19. Jaing TL, Hwang B, Lu JH, Hsieh KS, Meng CC. Percutaneous balloon valvuloplasty in severe pulmonary valvular stenosis. Angiology 1995; 46: 503-9.
20. Handoka NM, El-Eraky AZ. Predictors of Successful Pulmonary Balloon Valvuloplasty in Infants with Severe Pulmonary Valve Stenosis.Egypt Heart Mirror Journal 2007; 1: 68-74.
21. Qureshi SA, Redington AN, Wren C, Ostman-Smith I, Patel R, Gibbs JL,et al. Recommendations of the British Paediatric Cardiac Association for therapeutic cardiac catheterisation in congenital cardiac disease. Cardiol Young 2000; 10: 649-67.
22. Rao PS. Further observations on the effect of balloon size on the short term and intermediate term results of balloon dilatation of the pulmonary valve. Br Heart J 1988; 60: 507-11.
23. McCrindle BW. Independent predictors of long-term results after balloon pulmonary valvuloplasty.Valvuloplasty and Angioplasty of Congenital Anomalies (VACA) Registry Investigators. Circulation 1994; 89: 1751-9.


NEWS AND EVENTS

ANIMAL-BASED STUDIES


Research articles conducted on animals, will not be considered for processing or publication in the JPMA.

FOR REVIEWERS

ANNOUNCEMENT

SUPPLEMENT

INDEX

INSTRUCTIONS TO AUTHORS

COMMITTEE ON PUBLICATION ETHICS


This journal is a member of and subscribes to the principles of the Committee on Publication Ethics.

Copyrights © 2015 JPMA- All rights reserved
Powered by: PakCyber